LABOMAR TEKNOLOJİK ÇÖZÜMLER

Tel : 0216 417 70 01 / Fax : 0216  417 33 50

 
 

www.labomar.net   /  webinfo@labomar.net  

 

 

 

 
 

Dissolved Oxygen Meters: HI 9143 - HI 9145 Extended Range Dissolved Oxygen Meters

  • Rugged and waterproof, ideal for outdoor applications
  • Automatic calibration
  • Extended D.O. range

Measurements in the field can subject instrumentation to the inclemency of weather. Cold, rain, snow, dust and humidity, associated with use in the field, can cause condensation which may damage a meter, rapidly deteriorating its performance and life. HI 9143 and HI 9145 are designed for harsh outdoor applications, especially waste water treatment and fish farming. Simple on-site calibration requires no chemical solutions. Just expose the probe to air and press the CAL button. In a few minutes, the meter is calibrated and ready to use. Auto-calibration eliminates the need to use a screwdriver or other devices to adjust the reading. The new and improved HI 9143 and HI 9145 measure and display O2 from 0 to 300%, 0 to 45 mg/L and temperature from 0 to 50 °C. In addition to the features of HI 9145, HI 9143 also supplies the user with compensation of salinity and altitude factors. Both meters compensate for the temperature effect.

HI 9143 is supplied complete with D.O. probe HI 76407/4 with 4 m (13) cable, 2 spare membranes, electrolyte solution HI 7041S (30 mL), batteries, rugged carrying case and instructions. HI 9145 is supplied complete with D.O. probe HI 76407/4 with 4 m (13) cable, 2 spare membranes, electrolyte solution HI 7041S (30 mL), batteries, rugged carrying case and instructions.

 
 

Available Accessories:
HI 7041S | Refilling electrolyte solution (30 mL)
HI 710005 | 110V to 12VDC adapter
HI 710006 | 220V to 12VDC adapter
HI 76407/4 | D.O. probe with 4 m (13') cable
HI 76407/10 | D.O. probe with 10 m (33') cable
HI 76407/20 | D.O. probe with 20 m (67') cable
HI 76407A/P | Pack of 5 replacement membranes
HI 721317 | Rugged carrying case
 
Specifications:  
Range Oxygen
  0.00 to 45.00 mg/L
Range % Saturation Oxygen
  0.0 to 300.0%
Range Temperature
  0.0 to 50.0 °C
Resolution Oxygen
  0.01 mg/L
Resolution % Saturation Oxygen
  0.1%
Resolution Temperature
  0.1°C
Accuracy (@20°C/68°F)
  ±1.5% Full Scale (Oxygen); ±0.5°C (Temp)
Calibration
  Automatic in saturated air
Temp. Compensation
  Automatic from 32 to 122°F (0 to 50°C)
Auto Altitude Compensation: HI 9143
  0 to 1900m (0 to 6230')
Resolution: HI 9143
  100 meter
Auto Salinity Compensation: HI 9143
  0 to 40 g/L (ppt)
Resolution: HI 9143
  1 g/L (ppt)
Probe
  HI 76407/4 Polarographic D.O. probe
Battery Type / Life
  4 X 1.5 Volt AA / Power plug for 12VDC supply/ approx. 200 hours of continuous use
Environment
  32 to 122°F (0 to 50°C)
Dimensions
  7.7 x 3.1 x 2.4" (196 X 80 X 60mm)
Weight
  3.1 lb (1.4 Kg)
 

 Dissolved Oxygen Meters: HI 2400 Extended Range Bench Dissolved Oxygen Meter

  • Direct measurements in ppm or %
  • Logging of up to 8000 samples divided into 99 lots
  • Automatic calibration for D.O. and temperature in seconds with automatic value recognition

For Dissolved Oxygen measurements, HANNA instruments® presents the new HI 2400, an essential instrument for testing pharmaceutical and food products, as well as monitoring in water treatment plants. The user can choose to display D.O. readings in mg/L O2 or % of saturation. This meter can be used in any laboratory and for any type of water, as it allows measurement compensation both for altitude and salinity factors. The automatic logging interval can be set to perform analysis and store data into the non-volatile internal memory. All logged data can be downloaded to your PC through an RS232 serial port. Memory can store up to 8000 samples. HI 2400 features an automatic calibration procedure, at 1 or 2 points (at 0 and 100% of O2 saturation). The probe supplied with the meter is provided with a polarographic sensor, which measures the current generated by the reaction of O2 with Ag.

HI 2400 is supplied complete with HI 76407/2 DO probe with 2 m (6.6') cable, 2 spare membranes, HI 7041S electrolyte solution (30 mL), 12 Vdc power adapter and instructions.

 
 

Available Accessories:
HI 7040M | Zero oxygen solution (230 mL)
HI 7040L | Zero oxygen solution (500 mL)
HI 710005 | 110V to 12VDC adapter
HI 710006 | 220V to 12VDC adapter
HI 76407/10 | D.O. probe with 10 m (33') cable
HI 76407/20 | D.O. probe with 20 m (67') cable
HI 76407A/P | Pack of 5 replacement membranes
HI 92000 | Windows® compatible software
HI 920010 | PC connection cable
 
Specifications:  
Range
  0.00 to 45.00 mg/L O2; 0.0 to 300.0 % Saturation O2; 0.0 to 50.0 % Saturation °C
Resolution
  0.01 mg/L O2; 0.1 % Saturation
Accuracy (@20°C/68°F)
  ±1.5% of full scale mg/L O2; ±0.5 °C
Typical EMC Deviation
  ±1.5% of full scale mg/L O2; ±0.5 °C
Calibration
  Automatic single or double point at 0% and 100% (in air) mg/L O2; Automatic single or double point at 0.0°C and 50.0°C
Altitude Compensation
  0 to 1,900 m (6,230') with 100 m resolution
Salinity Compensation
  0 to 40 g/L with 1 g/L resolution
Temp. Compensation
  32 to 122°F (0.0 to 50.0°C)
Probe
  HI 76407/2 with 6.6' (2 m) cable
Logging Interval
  1, 15, 30 seconds or 1, 2, 5, 15, 30, 60, 120 & 180 minutes
Computer Interface
  RS 232C (opto-isolated)
Battery Type / Life
  12VDC (CSA, UL & CE approved)
Environment
  32 to 122°F (0 to 50°C); RH 95%
Dimensions
  9.1 x 6.7 x 2.7" (230 x 170 x 70 mm)
Weight
  2.2 lb (1 kg)

 

Dissolved Oxygen Meters: HI 2400Extended Range Bench Dissolved Oxygen Meter

  • Direct measurements in ppm or %
  • Logging of up to 8000 samples divided into 99 lots
  • Automatic calibration for D.O. and temperature in seconds with automatic value recognition

For Dissolved Oxygen measurements, HANNA instruments® presents the new HI 2400, an essential instrument for testing pharmaceutical and food products, as well as monitoring in water treatment plants. The user can choose to display D.O. readings in mg/L O2 or % of saturation. This meter can be used in any laboratory and for any type of water, as it allows measurement compensation both for altitude and salinity factors. The automatic logging interval can be set to perform analysis and store data into the non-volatile internal memory. All logged data can be downloaded to your PC through an RS232 serial port. Memory can store up to 8000 samples. HI 2400 features an automatic calibration procedure, at 1 or 2 points (at 0 and 100% of O2 saturation). The probe supplied with the meter is provided with a polarographic sensor, which measures the current generated by the reaction of O2 with Ag.

HI 2400 is supplied complete with HI 76407/2 DO probe with 2 m (6.6') cable, 2 spare membranes, HI 7041S electrolyte solution (30 mL), 12 Vdc power adapter and instructions.

 
 

Available Accessories:
HI 7040M | Zero oxygen solution (230 mL)
HI 7040L | Zero oxygen solution (500 mL)
HI 710005 | 110V to 12VDC adapter
HI 710006 | 220V to 12VDC adapter
HI 76407/10 | D.O. probe with 10 m (33') cable
HI 76407/20 | D.O. probe with 20 m (67') cable
HI 76407A/P | Pack of 5 replacement membranes
HI 92000 | Windows® compatible software
HI 920010 | PC connection cable
 
Specifications:  
Range
  0.00 to 45.00 mg/L O2; 0.0 to 300.0 % Saturation O2; 0.0 to 50.0 % Saturation °C
Resolution
  0.01 mg/L O2; 0.1 % Saturation
Accuracy (@20°C/68°F)
  ±1.5% of full scale mg/L O2; ±0.5 °C
Typical EMC Deviation
  ±1.5% of full scale mg/L O2; ±0.5 °C
Calibration
  Automatic single or double point at 0% and 100% (in air) mg/L O2; Automatic single or double point at 0.0°C and 50.0°C
Altitude Compensation
  0 to 1,900 m (6,230') with 100 m resolution
Salinity Compensation
  0 to 40 g/L with 1 g/L resolution
Temp. Compensation
  32 to 122°F (0.0 to 50.0°C)
Probe
  HI 76407/2 with 6.6' (2 m) cable
Logging Interval
  1, 15, 30 seconds or 1, 2, 5, 15, 30, 60, 120 & 180 minutes
Computer Interface
  RS 232C (opto-isolated)
Battery Type / Life
  12VDC (CSA, UL & CE approved)
Environment
  32 to 122°F (0 to 50°C); RH 95%
Dimensions
  9.1 x 6.7 x 2.7" (230 x 170 x 70 mm)
Weight
  2.2 lb (1 kg)