HI 98230 • HI 98240 Portable,
Printing and Logging pH/ORP/°C Meters with GLP
- Built-in Printer. - Uses plain
paper to print hard copies of data.
- "Intelligent" Electrode
Technology. - Allows the electrode to memorize the calibration
date and data performed with a specific instrument.
- Extended pH Range. - From -4 to
19.99 pH
- Dual Level Backlit LCD -
Unsurpassed visibility of measurements.
|
The top-of-the-range
HI 98230 and HI 98240 portable pH meters offer all the
features of HI 98140 and HI 98150 with the
advantage of having a built-in plain paper printer. Featuring GLP (Good
Laboratory Practice) capabilities, these instruments incorporate Hanna’s
latest “intelligent” electrode technology. These pH electrodes
incorporate a chip which memorizes the calibration data performed with a
specific instrument. The meter and electrode recognize each other and you
can calibrate a series of electrodes for specific applications and swap
them without needing to recalibrate the meter! This new technology allows
the operator to optimize time and efficiency with unsurpassed safety. It
also avoids erroneous measurements due to the pH electrode having been
substituted after calibration. This advanced series of electrodes also
incorporates a temperature sensor, eliminating the need for an additional
temperature probe. HI 98230 and HI 98240 allow the user to
print a permanent undisputable record of the measurement data at any time
for record keeping, troubleshooting or compliance purposes. The printing
interval is user selectable from 1 to 180 minutes. When the logging mode
is activated the instrument stores the measurement data which can be
printed directly or at a later date. The stored data (up to 500
measurements) can also be downloaded to a PC with a standard RS232 serial
connection and Hanna’s infrared adapter and Windows® compatible
software. HI 98230 and HI 98240 also feature a dual-level
backlit LCD which facilitates operation in dimly lit areas. Graphic
symbols and messages on the LCD guide the user through the measurement
procedures as well as during calibration. The pH range has been extended
to -4.00 to 19.99 pH and pH calibration is automatic at one or two points
with five stored buffer values (pH 4.01, 6.86, 7.01, 9.18 and 10.01). The
meter also allows the operator to set a calibration max. time that will
alert the user when recalibration is necessary. Access to the GLP data is
password protected for added safety. All pH measurements are automatically
compensated for the effects of temperature and a stability indicator will
appear on the display when the pH measurement is stable. HI 98230
and HI 98240 are each supplied with HI 1618D intelligent
electrode with built-in temp. sensor, DIN, 1 m (3.3’) cable, 12VDC
adapter, pH 4.01/7.01 buffer solutions, cleaning sol., 5 rolls of printer
paper, batteries, hard carrying case
|
Specifications:
|
|
HI
98230 |
HI
98240 |
Range: |
pH |
-4.00
to 19.99 |
-4.00
to 19.99 |
|
mV(ISE) |
— |
±400.0 |
|
mV(ORP) |
— |
±2000 |
|
°C |
-10.0
to 120.0 |
-10.0
to 120.0 |
Resolution: |
pH |
0.01 |
0.01 |
|
mV(ISE) |
— |
0.1
(-400 to +400); 1 |
|
mV(ORP) |
— |
0.1
(-400 to +400); 1 |
|
°C |
0.1 |
0.1 |
Accuracy: |
pH |
±0.01 |
±0.01 |
(@20°C/68°F) |
mV(ISE) |
— |
±0.5
(-400 to +400); ±2 |
|
mV(ORP) |
— |
±0.5
(-400 to +400); ±2 |
|
°C |
±0.5
(0 to 70); 1 |
±0.4
(0 to 70); 1 |
Typical |
pH |
±0.03 |
±0.03 |
EMC
Deviation: |
mV(ISE) |
— |
±1
(above ±400); ±2 |
|
mV(ORP) |
— |
±1
(above ±400); ±2 |
|
°C |
±0.8 |
±0.8 |
pH
Calibration: |
|
Automatic
1 or 2 points with 5 memorized buffers
(pH 4.01, 6.86, 7.01, 9.18 or 0.01)
|
mV
Calibration: |
|
— |
Automatic
2 points at 0, 350 mV
or 3 points at 0, 350, 1900 mV |
Temp.Compensation: |
|
Automatic
from -10 to 120°C (14 to 248°F) |
Electrode: |
|
HI
1618D intelligent electrode with
temp sensor, DIN, 1 m (3.3’) cable (included) |
Computer
Interface: |
|
HI
9200 (infrared interface)
connected to serial
RS 232 port
|
Printing/Logging
Interval: |
|
Selectable
from
1, 2, 5, 10, 15, 30, 60,
120 and 180 minutes
|
Input
Impedance: |
|
10¹²ohm |
Power
Supply: |
|
4
batteries 1.5V AA/300 hours approx.
continuous use
(without backlight) or 12VDC adapter (incl.)
|
Environment: |
|
32
to 122°F (0 to 50°C); RH 95% |
Dimensions: |
|
220x82x66
mm (8.6x3.2x2.6”) |
Weight: |
|
500
g (17.6 oz.) |
|
| |