LABOMAR TEKNOLOJİK ÇÖZÜMLER

Tel : 0216 417 70 01 / Fax : 0216 417 33 50

 
 

www.labomar.net   /  webinfo@labomar.net  

 

 

 

  ProfLine Oxi 197i
Range/Resolution O2 concentration: 0.00 to 19.99 mg/l, 0.0 to 90.0 mg/l
O2 saturation: 0.0 to 199.9%, 0 to 600%
Temperature: -5.0 to ±50.0°C
Accuracy (± 1 digit) O2 concentration: ±0.5% of value
O2 saturation: +0.5% of value
Temperature: +0.1 K
Air Pressure Compensation automatic with built-in pressure sensor (500 to 1100 hPa)
Salinity Correction automatic from 0.0 to 70.0, adjustable via display
Calibration OxiCal® rapid calibration in OxiCal®-SL

Ordering Information
Portable Oxygen Meter Order No.
ProfiLine Oxi 197i   Robust, waterproof, submersible oxygen meter 3B30-010

Model

Oxi 315i

Oxi 330i

Oxi 340i

Measuring range/ resolution O2 concentration 0.00 … 19.99 mg/l (19.9 mg/l*)  0.0 … 90.0 mg/l (90 mg/l*) O2 concentration 0.00 … 19.99 mg/l (19.9 mg/l*)  0.0 … 90.0 mg/l (90 mg/l*)
O2 saturation 0.0 … 199.9% (199%*)   0 ... 600% O2 saturation 0.0 … 199.9% (199%*)    0 … 600%
. O2 part. press. 0,0 … 199.9 mbar (199 mbar*)    0 ... 1250 mbar
temperature -5.0 … +50.0 °C temperature -5.0 … +50.0 °C
*if DurOx oxygen sensor is used *if DurOx oxygen sensor is used
Accuracy (±1 digit) O2 concentration ±0.5% of value O2 concentration ±0.5% of value
O2 saturation ±0.5% of value O2 saturation ±0.5% of value
. O2 part. press. ±0.5% of value
temperature ±0.1 K temperature ±0.1 K
Temperature compensation 2% at 0 ... +40 °C
Air pressure compensation automatic with built-in pressure sensor (500 ... 1100 hPa)
Salinity correction 0 or 35.0 solid SAL automatic from 0.0 ... 70.0, adjustable via display
Display LCD display 60 x 45 mm visible area, simultaneous display of measurement, temperature, special characters
Sensor evaluation depends on calibration results, shown on display
Calibration alarm 1 ... 999 days, adjustable -
AutoRead for reproducibility better than 0.05 mg/l
Calibration OxiCal® rapid calibration in OxiCal®-SL or OxiCal®-D
Calibration protocol - - GLP-conform calibration protocol; last protocol stored in memory
Data storage - 500 data sets (measurement, temperature, date/time, identity no.); time-controlled storage at intervals 5 s ... 60 min, non-volatile data storage (even when changing batteries)
Data output via display via display or interface
Interface - -

self-configuring analog output 0 ... 2 V corresponding to display, or RS 232 bi-directional interface

Operating time

2000 hours with normal 4 x 1.5 V batteries continuous operation from mains with AutoSwitchOff after 1 h (except logger operation)

 

supply; 600 hours with battery operation with  AutoSwitchOff after 1 h(except logger operation)
Protection type IP 66 and IP 67 to IEC 529
Dimensions 172 x 80 x 37 (H x B x D)
Weight approx. 300 g
Certified to CE, TÜV/GS, UL, CUL
Warranty 3 years on the instrument